Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT18220870636654301049466531

LightSpeed 3.X (Ultra, Plus, QX/i) - GRE

Symptom

Cradle will not move in or out of the Gantry. No c-pulse is being detected from longitudal encoder Cradle will not move in or out of the Gantr . No c-pulse is being detected from longitudal encoder y

Error message

Error: 182208 Alignment error between Encoder and Potentiometer. Potentioment Position 70636 (.01 mm) Encoder Position 65430 (.01 mm) Tolerance 2540 (.01 mm) See the LD for the complete error code Error: 182208 Alignment error between Encoder and Potentiometer. Tue Aug 5 15:05:36 1997 Suite: PCT2 Host: ETC Proc:Cradle Error: 182208 File: cmcEncoderRef.c 1.8 8/27/ Method: No Method Line: 152 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 10494/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:05:35:935 Log Series: 27 HW error: Encoder C Pulse Alignment error with current Potentiometer reading. Current Position 66531 (.01 mm). Encoder Alignment Minimum (1456) and Maximum (1658) Potientiometer Window based on Last Potentiometer Reading (1658).

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2271256 SIGNAL HARNESS FOR H2 TABLE See Long Description for part details 2271256 SIGNAL HARNESS FOR H2 TAB Part #: 2271256 Part Type: GP CONSUMABLES Part Weight: 1.09 LBR Repairable: No Returnable: No Hazardous Material: N Description: SIGNAL HARNESS FOR H2 TABLE FRU: Yes Status: Orderable Comments: FLS - THIS IS SIGNAL HARNESS USED ON LIGHTSPEED H2/FRU CODE GPCP/ FOUND NO STOCK/PDR ALREADY IN PLACE/MS6255 01/19/02 ASIA SOURCED PART
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required