Tech received a system level error pop up during fastcals.

Symptom

Tech received a system level error pop up during fastcals. Tech received a Critical software failure pop up during fastcals

Error message

Error: 244634 Subsystem State; DCB:up, CCB:up, TCB:up, JEDI:up, --- TGP/ORP:down, TGP/ORP:DOWN See Long Description Sun Apr 13 14:45:14 2008 Error: 244634 Host: lawct2 Process: fwmgr File: fwmgr_utils.c Line: 664 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem State; DCB:up CCB:up TCB:up JEDI:up TGP/ORP:down .

Error: 260116313 Table Gantry Processor - Host (OC) LAN Communication Watchdog Timed out. see LD Tue May 16 22:03:29 2006 Error: 260116313 Host: tgp Process: tgp_subsys File: subsys_mgr.c Line: 349 Running: System Reset Tue May 16 21:37:53 2006 Host : Table Gantry Processor Ermes # : 260116313 Exception Class : Safe State Severity : Secondary File : subsys_mgr.c Line# : 349 Function : SYSTEM COMMUNICATIONS : Control Network Scan Type : None/Unknown Scan: 0/0/0 Table Gantry Processor - Host (OC) LAN Communication Watchdog Timed out.

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.