GE Healthcare · CTLightSpeed210000445210000447210000426
LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry.
Symptom
Unable to display any images; exam browser listing in image works is empty and fails to refresh.
Error message
Error: 210000445 Error: saveItem() failed with status X. See Long Description for complete error. Fri Sep 30 14:59:58 2005 Error: 210000445 Host: ct99 Process: dms File: DMFInstall.cxx Line: 236 Error: saveItem() failed with status -1 error message TdbExceptionNotImplem ented- for image path /usr/g/sdc_image_pool/reserved/IMSsecBin16/E613U0_S 2U0_I25U0.1128106798_69694 Thu Jul 30 08:55:26 2009 Error: 210000445 Host: H0140CT1 Process: dms File: DMFInstall.cxx Line: 450 Error: saveItem() failed with status 0 error message DMFCreateSession() failed for image path /usr/g/sdc_image_pool/reserved/IMSse7Q2TVf/E1009U0_S1U0_I1U0.1248958526_266351 Error: 210000447 unable to connect to session dms_session with error code -2 See long description for details Tue Jan 29 06:51:29 2008 Error: 210000447 Host: gevct2 Process: dms File: DMFInstall.cxx Line: 229 Error: unable to connect to session dms_session with error code -2 error message Session Not Connected Error: 210000426 Unable to install image Exxxxxxxxx with status -1 See Long Description for complete error. Fri Sep 30 14:59:58 2005 Error: 210000426 Host: ct99 Process: dms File: InstallImage.cxx Line: 551 Unable to install image E613S2I25
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 08BW44.1 Software for all BrightSpeed Systems with AIO Console and all RT16 Systems with GOC4 Console See Long Description 08BW44.1 Software is used on all BrightSpeed Systems with AIO Consoles and all RT16 Systems with GOC4 Consoles. Software collector called “08BW44.X” p/n 5328867 contains the following items: 5323837 - LightSpeed Xtream Console Operating System Version - GEMS Linux 6.3.8 5323838 - LightSpeed Applications Software DVD Version 08BW44.X 5328828 - Xtream Serial-Over-LAN Service Software Version - 2.0 Supports: Westville or Jarrell DARC/VDARC 5323839 - Class C 5323840 - Class M NOTE: Class C/M install CD are not shipped with system. NOTE: The Xtream Serial-Over-Lan Service Software is not used during the LFC unless instructed via a pop-up message box "SOL Service - ...run the Serial Over LAN Service CD...". This CD is loaded directly into the DARC Node. Then press the reset (or cycle power) at the front panel of the DARC Node as required. This CD will auto-eject when the SOL load process is complete. NOTE: The Xtream Serial-Over-Lan CD is not used for All-In-One Console.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required