GE Healthcare · CTBrightSpeed290009023290012004
BrightSpeed HPower, Delight, Velocity or MERI HPower
Symptom
Recon self test failure on boot up; scanning would be allowed but no reconstruction of images. Recon self test failure on boot up; scanning would be allowed but no reconstruction of images
Error message
Error: 290009023 Resetting IG (1,2,3) because it has reached max consecutive retries: 4 See Long Description for complete error. Error: 290009023 Resetting IG (1,2,3) because it has reached max consecutive retries: 4 Sun May 14 13:40:24 2006 Error: 290009023 Host: darc Process: recon_control File: IG_interface_list.cxx Line: 1272 Resetting ig2 because it has reached max consecutive retries: 4 Note: This could be IG1, IG2, IG3 Error: 290012004 RAC hardware failure: error Making image See Long Discription for Details Mon Aug 2 08:37:19 2010 Error: 290012004 Host: ig1 Process: image_generation File: RacInterface.cpp Line: 780 RAC hardware failure: error making image
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5147443-2 Jarrell IG for NON VCT systems (replaces 5147443 ). See long description for details Part #: 5147443-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Jarrell GRE IG with RAC rev.D or Higher FRU: Yes Status: Orderable Comments: Per ECR 2040301 (Part of Jarrell NPI), 212008898, 3/8/07. **FE Assist** applies to CT but not PET systems. Must order software release 07MW11.10, p/n 5220443 (deployed by FMI 25389), and perform a LFC to support this FRU if system is not currently running sw rls 06MW29.7 or 07MW11.10 (8/16/07 212001261) 212008041 12/03/2007 15:23 GMT Part #: 5147443 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Computer - Image Generation, 2nd Generation FRU: Yes Status: Orderable Comments: Per Jarrell RSPL, 7/19/06, ms11892. ** FE Assist ** If ordering this Jarrell IG to replace a Westville IG 2362872 must also order a 5127222-2 Shielded Network cable and replace the existing one on your system Plus, must order software release 06MW29.7, p/n Part #: 5193989 and perform a LFC. (12/8/05 212001261
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required