Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26010107150018260101008

LightSpeed 5.X (60kW)

Symptom

Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .

Error message

Error: 260101071 The DCB has reset unexpectedly See Long Description for full error Error: 260101071 Host: orp Process: orp_subsys File: rcib_node_init.c Line: 376 Wed Jan 24 11:25:20 2007 Host : On-Board Rotating Processor Ermes # : 260101071 Exception Class : Secondary Severity : Pri/Soft File : rcib_node_init.c Line# : 376 Function : No System Function Reported Scan Type : None/Unknown Scan: 50018/1/0 The DCB has reset unexpectedly. The ORP has successfully reconnected to sub-system. Reason for reset as reported by CPU: External signal source. CPU reset register: 0x80 Node status: 0x382 State flags: 0x9800e82

Error: 260101008 The ORP failed to connect with the DCB sub-system in the allotted time. See LD Wed Jul 12 06:02:29 2006 Error: 260101008 Host: orp Process: orp_subsys File: rcib_node_init.c Line: 110 Wed Jul 12 06:02:30 2006 Host : On-Board Rotating Processor Ermes # : 260101008 Exception Class : Soft Severity : Pri/Soft File : rcib_node_init.c Line# : 110 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 The ORP failed to connect with the DCB sub-system in the allotted time. Check to see if subsystem is alive. Timeout value: 40000mS

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2361853-2 GDAS DCB 16-SLICE replaces 2361853 See LD for more Info 2361853-2 GDAS DCB 16-SLICE replaces 2361853 Part #: 2361853-2 Part Type: GP CONSUMABLES Part Weight: -59999.40 LBR Repairable: No Returnable: No Hazardous Material: N Description: GDAS DCB FRU: Yes Status: Orderable FLS - GDAS DCB (16-SLICE) USED ON: LIGHTSPEED PRO 16 GDAS ADDED PER GDAS RSPL MS13452 This board has been replaced by 2361853-2, FE NOTE: Site MUST have M4FMI completed before this board will function properly, ms13452
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.