GE Healthcare · CTLightSpeed2600073531095852101260007400109585207637776260007771
LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry.
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260007353 Scan ABORT -- DAS Has Detected An Error._ Scan Setup Errors Prevent The Data Acquisition System From Being Able To Scan. (See Previous Log Entries.) 1095852101 0 1 Wed Sep 22 11:21:41 2004 260007353 orp orp_subsys setuprx.c 169 Wed Sep 22 11:21:39 2004 Host : Das Ermes # : 260007353 Exception Class : Abort Severity : Primary File : setuprx.c Line# : 169 Function : No System Function Reported Scan Type : None/Unknown Scan: 26/2/0 Scan ABORT -- DAS Has Detected An Error._ Scan Setup Errors Prevent The Data Acquisition System From Being Able To Scan. (See Previous Log Entries.) 1095852101 0 1 Wed Sep 22 11:21:41 2004 260007353 orp orp_subsys setuprx.c 169 Wed Sep 22 11:21:39 2004 Host : Das Ermes # : 260007353 Exception Class : Abort Severity : Primary File : setuprx.c Line# : 169 Function : No System Function Reported Scan Type : None/Unknown Scan: 26/2/0 Scan ABORT -- DAS Has Detected An Error._ Scan Setup Errors Prevent The Data Acquisition System From Being Able To Scan. (See Previous Log Entries.) Error: 260007400 Command 106 to Converter Manager FAILED See LD for complete error Error: 260007400 Command 106 to Converter Manager FAILED 1095852076 1 1 Wed Sep 22 1
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2404393 GDAS-16 CONVERTOR CARD, replaces 2364354. See Long Description for details. 2404393 GDAS-16 CONVERTER CARD, replaces 2364354 Part #: 2404393 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: 16 SLICE GDAS CONVERTER CARD FRU, A CONVERTER CARD IN AN ESD BAG IN A BOX WITH A PART NUMBER LABEL FRU: Yes Status: Orderable Comments: FLS - 16 SLICE GDAS CONVERTER CARD USED ON: LIGHTSPEED PRO 16 FRU GDAS ADDED PER GDAS RSPL MS13452 040315 RISK: SITE DOWN FREQUENCY: HIGH NOTE: THERE ARE 96 OF THESE CARDS PER SYSTEM, BUT SHOULD NOT NEED TO REPLACE ALL AT ONETIME, MS13452 THIS PART CONTAINS ONE GDAS CONVERTER CARD 2364354 WITH FRU PACKAGING, SUPPLIED BY GEMS IT. MS13452 PDR 8471.80.9000 IM05 UPDATE AWAITING MM (212001261 4/1/06) For 4/8 slice G Das which uses a thick connector must order 2364354-2
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.