Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT26000314113155260003158

LightSpeed 3.X (Ultra, Plus, QX/i) - GRE

Symptom

Scanning aborts with collimator filter homing problem. Scanning aborts with collimator filter homing problem

Error message

Error: 260003141 Filter homing problem. See the Long Description for the full error code Fri Mar 4 21:22:12 2005 Host : Collimator Ermes # : 260003141 Exception Class : Abort Severity : Primary File : filter_home.c Line# : 147 Function : No System Function Reported Scan Type : None/Unknown Scan: 13155/1/0 Filter homing problem.

Error: 260003158 The Filter Home Flag Has Moved Since The Last Reference See long description for full error Fri Feb 17 17:56:41 2006 Error: 260003158 Host: HSC Process: CCB Firmware File: filter_home.c Line: 193 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 8650/2/0 Type: None/Unknown Exception Level: Sec/Soft Time: 17:56:41:390 Log Series: 564 The Filter Home Flag Has Moved Since The Last Reference. Possible Causes: Filter Home Switch, Filter Encoder, Filter Encoder Set Screw. Difference: 360 degrees

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5331736 - Helios Filter Asselmbly FRU (replaces 2291958) Used on Helios 3 and 4 Collimators. See Long Description for more Details Part #: 5331736 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Helios Filter Asselmbly FRU FRU: Yes Status: OrderableComments: Per ECR 2082045, 212008898, 6/16/09. Part #: 2126674 Part Type: GP CONSUMABLES Part Weight: -144548.55 LBR Repairable: No Returnable: No Hazardous Material: N Description: FILTER ASSEMBLY, HELIOS COLLIMATOR FRU: NO Status: Non Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues