GE Healthcare · CTLightSpeed200002368
LightSpeed RT16 & RT16 Xtra and Vision
Symptom
System will not boot to or stay in applications. System will not boot or stay in applications
Error message
Error: 200002368 A set of processes are deadlocked waiting for each other to start. This deadlock has caused the system to shutdown! See Long Description for full error. Mon Oct 9 07:55:18 2006 Error: 200002368 Host: ct01 Process: cupMonitor File: ssProcStart.c Line: 997 A set of processes are deadlocked waiting for each other to start. "afQueueMgr, examRxDisplayMgr, sfm_archive, scanmgr, scanfilemgr, headermgr, coordinator, rtsmgr, reconSelftest, queuemgr, procICE, accs, vavQueueMgr, stcIf, sdm, scanRx, retroRecon, reconMgmt, readcfgServer, ppsproxy, piRxMgmt, mtran, examRxEnv, dms, visScript, ces, awStart, WLServer, ObjMgr" are the likely culprits. This deadlock has caused the system to shutdown!
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.