Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCTDiscovery290014005

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Pink Pop-up Scan Disk Array has encountered a hard drive failure. If another hard drive fails you will lose scan data. See Long Description: Pink Pop-up message is displaced to the customer on boot up. Scan Disk Array has encountered a hard drive failure. If another hard drive fails you will lose scan data. "Scan Disk Array has encountered a hard drive failure. The System is functional, but if another hard drive fails you will lose scan data. Please Contact GE Service to have the disk array repaired as soon as possible."

Error message

Error: 290014005 The scan disk array was started but a disk failure was detected. See Long Discription for Details Tue Sep 14 06:42:19 2010 Error: 290014005 Host: sccact1 Process: gre-verify-darc File: gre-verify-darc.cpp Line: 185 The scan disk array was started but a disk failure was detected. The logical RAID has been marked as degraded. Host: darcarray Disk: 3

Applies to

Models
LightSpeed, VCT, Discovery
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5308759 2U 12Bay SAS to SAS-SATA RAID See Long Discription for Details Part #: 5308759 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: 2U 12Bay SAS to SAS-SATA RAID Rack -RAID6 Capable- including the following components - Single controller with 512MB RAM, ASIC400, 4xSAS Host Ports, 1xSAS Expansion Port, Dual Hot-swappable 460W PS, Redundant Dual-speed Cooling Modules, an FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required