GE Healthcare · CTLightSpeedVCT260006605500002600068125046226000682318729
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
"Scan hardware Stopped Scan" with Axial Errors. "Scan hardware Stopped Scan" with Axial Errors
Error message
Error: 260006605 Could not send axial drive message in the allotted time. Axial serial link is down. See the LD for the full error code Error: 260006605 Could not send axial drive message in the allotted time. Tue Nov 16 14:20:36 2004 Error: 260006605 Host: STC Process: Axial File: scanbusDriver.c 1.5 Line: 523 Function: AXIAL CONTROL : axial communications Scan: 50000/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 14:20:37:325 Log Series: 70 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices, .. Error: 260006812 Gantry Failed to change to Commanded Velocity See the LD for the full error code Error: 260006812 Gantry Failed to change to Commanded Velocity Error: 260006812 Host: STC Process: Axial File: amc_exec.c Line: 3261 Running diagnostic: Axial2 Functional Function: AXIAL CONTROL : axial control Scan: 50462/11/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 13:32:33:890 Log Series: 249 Gantry Failed to change to Commanded Velocity of 4.0 secs/rev within 15.000 seconds. (AMC state:4 Cmd:0x159a0028 CurVel:4160 (.001 sec/rev)) Retry Rx to verify pr
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP, VCT, and HD750. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) See LD for details 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP and VCT. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) Part #: 5311936 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Shipping collector for Axial Encoder Assembly. FRU: Yes Status: Orderable Comments: Per ECR 2058601, 212008898, 10/8/08
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.