GE Healthcare · CTLightSpeedVCT26011826050068
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260118260 ( 80-1422 ) Overpress or Blown fuse Error: 260118260 ( 80-1422 )Overpress or Blown fuse Wed Aug 17 17:37:03 2005 Error: 260118260 Host: orp Process: orp_subsys File: gen_exec.c Line: 2161 Running: System Reset Wed Aug 17 17:36:57 2005 Host : Jedi Ermes # : 260118260 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2161 Function : No System Function Reported Scan Type : None/Unknown Scan: 50068/1/0 (80-1422)Overpress or Blown fuse. Z6,Z7= 0h,0h
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5125429 FUSE KIT for VCT systems See Long Description for details 5125429 FUSE KIT for VCT systems NOTICE: The information below came from a document that is under revision control and it is possibly out of date. Always check for the latest revision of the manual if necessary . # NUMBER QTY FRU DESCRIPTION/NAME 1 2142415-4 3 Yes PROTECTION FUSE, 5A, TIME DELAY, 600VAC 250VDC, BOARDS USE FUSE CLIPS 2 2142415-5 3 Yes PROTECTION FUSE, 8A, TIME DELAY, 600VAC 250VDC, BOARDS USE FUSE CLIPS 3 2142415-7 3 Yes PROTECTION FUSE, 15A, TIME DELAY, 600VAC 250VDC, BOARDS USE FUSE CLIPS 4 2142415-9 2 Yes PROTECTION FUSE, 25A, TIME DELAY, 600VAC 250VDC, CHASSIS MOUNTED-- CURRENT RATING HIGHER THAN FOR TYPICAL WITH FUSE CLIPS 5 2336517-2 2 Yes FUSE 25 AMPS 700 VOLTS 2.5M SECONDS 6 2351492 3 Yes 200A FUSE 7 2364059 2 Yes GLASS FUSE 8 2379651 2 Yes FUSE - 700 V, 200 A, FAST SEMI 9 46-170021P106 2 Yes FUSE 8 AMPS 250 VOLTS BUSSMANN ONLY 10 46-170021P75 2 Yes 015.000A 0250V 3AB SLOW UL BUSSMANN MDA 15 11 46-267217P20 4 Yes FUSE METRIC SLO BLO 2.500 A 250V 5X20MM GLASS LITTLEFUSE 218 02.5 12 46-267217P23 2 yes FUSE METRIC SLO BLO 5.000 A 250V 5X20MM GLASS LITTLEFUSE 218 005 13 5118644 1 Yes FAST ACTION 15 AMPS 250 VOLTS 0.01 SECONDS U0028FF 14 U8006FF 3 Yes FUSE 1.0A 15 U8008FF 3 Yes FUSE 2.0A 16 U8009FF 3 Yes FUSE 3.2A 17 2390269 3 Yes FUSE 2.5 AMPS 125 VOLTS 5105346-3 (replaces 5105346-2 and 5105346) RoHS Compliant Performix PRO 100 Smart Pump - VCT or HD See LD for details. 5105346-3 RoHS Compliant Performix PRO 100 Smart Pump - VCT or HD Part #: 5105346-3 Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: RoHS Compliant Performix PRO 100 Smart Pump - VCT or HD FRU: Yes Status: Orderable Old FRU Compatible Yes Comments Rohs- Old part can be used on new systems. Flag to yes mar 18 2011 212008958 Per ECR 2100751, 212008898, 4 Jan 2011. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5105346-3 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number ECR2100751 Part #: 5105346-2 VCT PUMP / ADAPTER / HOSE ASM. Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Performix PRO 100 VCT Pump, Hose Asm. and Mounting Bracket Assembly, With furnished mounting hardware Pump Support Bracket and U-bolt Assembly and Flush and Filled with Dielectric Oil. FRU: Yes Status: Orderable Comments: This part is replacement for 5105346. Do NOT use 5105346 to replace a 5105346-2, ms13452 060131
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required