GE Healthcare · CTLightSpeed260134805
LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry.
Symptom
Table does not respond to movement commands. Table does not respond to movement commands
Error message
Error: 260134805 IMS tape switch IN (rear) isn't present See Long Description for more Details Thu Nov 10 08:55:28 2005 Error: 260134805 Host: tgp Process: tgp_subsys File: gtcb_imc_isrfunc.cxx Line: 163 Thu Nov 10 08:55:28 2005 Host : Table Sub-System Ermes # : 260134805 Exception Class : Secondary Severity : Pri/Soft File : gtcb_imc_isrfunc.cxx Line# : 163 Function : No System Function Reported Scan Type : None/Unknown Scan: 405/2/0 IMS tape switch IN (rear) isn't present. Position is 84 um.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5164938 Tape Switch (208mm) used as IMS Tape Switch IN (rear) replaces 5127462-2 - used on GT2000/GT1700. See long description for details Part #: 5164938 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TAPE SWITH 4 ASSY Positioning GT FRU: Yes Status: Orderable Comments: added per to GT SPCN. source pole/planner code revised per to Global Table transfer Note: the little plastic screw for this part is U0087AA The Latest info is in this table. Switch Type Location Part Number (Latest) Switch Description in Block Diagram Tape Switch Part Number Description Switch Description in gesyslog Purpose Tape Switch TOP COVER FRONT ASSY 5127465-3 T-Sensor_Elev1 T-Sensor_Elev2 T-Sensor_Elev3 T-Sensor_Elev4 5133731 TAPE SWITCH 2 (185mm) Elevation touch switch1 (tape switches on underside of top front cover) Detection of clog in elevation down 5133732 TAPE SWITCH 3 (265mm) Elevation touch switch1 (tape switches on underside of top front cover) Detection of clog in elevation down IMS LEFT SIDE COVER 5127463-2 T-Sensor_IMS3 5133730 TAPE SWITCH 1 (100mm) IMS tape switch OUT (front) Detection of pinching between IMS and column in elevation down T-Sensor_Elev10 5133732 TAPE SWITCH 3 (265mm) Elevation touch switch4 (tape sw on underside of IMS side cover) Detection of clog in elevation down IMS REAR UPPER COVER 5127462-2 T-Sensor_IMS1 T-Sensor_IMS2 5164938 TAPE SWITCH 4 (208mm) IMS Tape Switch IN (rear) Detection of pinching between IMS and column in elevation up IMS RIGHT SIDE COVER 5127464-2 T-Sensor_IMS4 5133730 TAPE SWITCH 1 (100mm) IMS tape switch OUT (front) Detection of pinching between IMS and column in elevation down T-Sensor_Elev11 5133732 TAPE SWITCH 3 (265mm) Elevation touch switch4 (tape sw on underside of IMS side cover) Detection of clog in elevation down Limit Switch GT REAR TOUCH SWITCH 5125112 T-Sensor_Elev7 T-Sensor_Elev8 T-Sensor_Elev9 N/A N/A Elevation touch switch3 (switch on underside of top rear cover) Detection of clog in elevation down TOUCH SENSOR R ASSY 5127469 T-Sensor_Elev5 N/A N/A Elevation touch switch2 (switches on underside of top side cover) Detection of clog in elevation down TOUCH SENSOR L ASSY 5127470 T-Sensor_Elev6 N/A N/A Elevation touch switch2 (switches on underside of top side cover) Detection of clog in elevation down
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.