GE Healthcare · CTLightSpeed24462160333
LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry.
Symptom
Tech received a system level error pop up during fastcals. Tech received a Critical software failure pop up during fastcals
Error message
Error: 244621 Timed out while attempting to receive a message. See long description for full error. Wed Aug 22 07:19:18 2007 Error: 244621 Host: mcc1 Process: fwmgr File: fwmgr fw if.cxx Line: 99 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Did not receive reply from a request. Reply Event : 60333 Reply Source: rcib Error: Timed out while attempting to receive a message
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT