GE Healthcare · CTLightSpeed230023052244634
LightSpeed 5.X (60kW)
Symptom
Tech received a system level error pop up during fastcals. Tech received a Critical software failure pop up during fastcals
Error message
Error: 230023052 Failed in Interprocess Communication with Firmware. See long description for details Fri Apr 4 07:32:59 2008 Error: 230023052 Host: chhvct Process: dailyPrepRx File: DailyPrepShellMgr.c Line: 2618 Failed in Interprocess Communication with Firmware. Critical software failure. Cannot continue. Error: 244634 Subsystem State; DCB:up, CCB:up, TCB:up, JEDI:up, --- TGP/ORP:down, TGP/ORP:DOWN See Long Description Sun Apr 13 14:45:14 2008 Error: 244634 Host: lawct2 Process: fwmgr File: fwmgr_utils.c Line: 664 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem State; DCB:up CCB:up TCB:up JEDI:up TGP/ORP:down .
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required