Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed200001202200350158

BrightSpeed Select, Delight Select or MERI Select

Symptom

Gantry service balance does not run See Long description for additional details After system reboot, clicks 'Gantry Balance'. There isn't any GUI shown up. Then click 'Gantry Balance' at the second time. GUI shows up. After clicks 'Run', there isn't response coming up. Close the GUI,clicks 'Gantry Balance' at the third time. GUI shows up. After click 'Run', The error log shows it can't acquire resource 200001202 could not communicate with resurce management server within 10 seconds. 200350158 cannot aquire resurce required for test

Error message

Error: 200001202 Could not communicate with resource management server within 10 seconds See Long Description Fri Feb 26 07:17:15 2010 Error:200001202 Host: motct Process: ddc File: rm_process_request.c Line: 100 Could not communicate with resource management server within 10 seconds

200350158 - Can not acquire resource required for test.

Applies to

Models
BrightSpeed
Product
BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
08BW44.1 Software for all BrightSpeed Systems with AIO Console and all RT16 Systems with GOC4 Console See Long Description 08BW44.1 Software is used on all BrightSpeed Systems with AIO Consoles and all RT16 Systems with GOC4 Consoles. Software collector called “08BW44.X” p/n 5328867 contains the following items: 5323837 - LightSpeed Xtream Console Operating System Version - GEMS Linux 6.3.8 5323838 - LightSpeed Applications Software DVD Version 08BW44.X 5328828 - Xtream Serial-Over-LAN Service Software Version - 2.0 Supports: Westville or Jarrell DARC/VDARC 5323839 - Class C 5323840 - Class M NOTE: Class C/M install CD are not shipped with system. NOTE: The Xtream Serial-Over-Lan Service Software is not used during the LFC unless instructed via a pop-up message box "SOL Service - ...run the Serial Over LAN Service CD...". This CD is loaded directly into the DARC Node. Then press the reset (or cycle power) at the front panel of the DARC Node as required. This CD will auto-eject when the SOL load process is complete. NOTE: The Xtream Serial-Over-Lan CD is not used for All-In-One Console.
Area
CT

Full cause & step-by-step resolution