Pular para o conteúdo principal
GE Healthcare · CT

Sytec 1600i/1800i SYj Eu

Symptom

Unable to scan with XG errors

Error message

ER-117 XG Fail - contains information on all XG ER117 errors See Long Description ER-117 XG Fail ER–117 ER – 117 : XG Error 00000XYY Error Code X = _ 0: Test Mode 1: Stand–by 2: Preparation Acting 3: Generator Ready 5: Exposure 7: Transition Step to Stand–by ER–117 (continued) ER – 117 : XG Error 00000XYY Error Code YY = _ X01: No DC550V X02: 15V_Safe Error X03: Over Voltage Capacity X11: Rotor Controller Error X12: Rotor Acceleration Error X13: Rotor Feedback Test X14: Rotor Fault Detect Test X16: Low Current Error X18: Unbalance Error X21: Over Filament Current 021: Feedback Test X22: No Filament Current 022: Fault Detect Test X23: Bad Relay Return 023: Small Focus Test X24: No Filament Voltage 024: Large Focus Test X25: Filament Control Error 526: MA measure Error 031: Feedback Test 032: Fault Detector Test 033: Close Loop Test 531: MA Accuracy Error 532: KV Accuracy Error 533: Spit Error 534: Exposure Backup Timer Error 535: Anode Over mA 536: Cathode Over mA 537: Anode Over kV 538: Cathode Over kV 539: Anode Shoot Through 540: Cathode Shoot Through 541: Anode Over Current 542: Cathode Over Current X43: HV ON with Error 544: KV Unbalanced Error X48: Generator Safety Loop Open X

Applies to

Product
Sytec 1600i/1800i SYj Eu Sytec 1600i/1800i SYj Eu Sytec 1600i/1800i SYj Eu Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2152636 Rotor cable XG-Rotor Sytec (eqvt P9210BK) see long description 2152636 Rotor cable XG-Rotor Sytec (eqvt P9210BK) Part #: 2152636 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: P9210BK FAN CABLE ASSY FRU: Yes Status: Orderable Parts Offered as: Current: 2152636 Harvested: Not Available Exchange: Not Available Comments: FLS - ROTOR (XG-TUBE) CABLE USED ON GANTRY OF CT SYTEC 1800I SYJ XG..C.BRICKLER 6/2/97
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues