GE Healthcare · CTLightSpeedVCT2446531163411915
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Attention The detector is not at the proper temperature. see long description for more detail. Attention The detector is not at the proper temperature. If you wish to proceed with this scan select the "OK" button, however IQ may be degraded including image artifacts. Please contact GE Service
Error message
Error: 244653 Host Querying Detector Temperature Status See LD for complete error Sun Jun 25 13:08:53 2006 Error: 244653 Host: gect02 Process: fwmgr File: fwmgr_events.c Line: 11634 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Host Querying Detector Temperature Status: StateFlag:2 curTe mp:3180, leftTemp:3620 rightTemp:3730 Time:0 Fri Feb 25 07:51:11 2011 Error: 244653 Host: MT_CT2 Process: fwmgr File: fwmgr_events.c Line: 11915 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Host Querying Detector Temperature Status: StateFlag:5 curTemp: 0 , leftTemp: 0 rightTemp: 0 Time:0 <--- All temperatures reporting zero - Check for Blown F6 on Fuse Board
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5314357 or equivalent 5174115 Routing Board Harness with filter - used on VCT and HD (5314357) See Long Description for details Part #: 5314357 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 KGM Repairable: No Returnable: Yes Hazardous Material: N Description: Harness-Routing board HD FRU: Yes Status: Orderable ====================== Parts Offered as:Current: 5314357 Harvested: Not Available Exchange: Not Available Comments: Per HDVCT Options RSPL, 212008898, 5/5/09. Old FRU Compatible Yes Comments Per Titan RSPL, 212008898, 5/3/2010. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands No Use Demand History Yes Ultimate Prime 5174115 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number BKS32447 Part #: 5174115 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Routing Board Harness with filter FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution