Revolution XRD-2X (Lightning) exposure stops intermittently Spit Max error Customer call service for some time…

Revolution XRD-2X (Lightning) exposure stops intermittently Spit Max error

Symptom

Revolution XRD-2X (Lightning) exposure stops intermittently Spit Max error Customer call service for some time the system can't do any exposure when system report " Tank Thermal sensor to high", after reboot, the lighting DR exposure stops intermittently the error is " generator/tube spits"

Error message

Spit Max error, Generator temperature too high, kV max detected, ILR max current detected China support Lindewei: " Generator error 60-0310: kV max detected" KV reach 160KVP during exposure was detected. The error also cause by KV control board defect or HV tank arcing during exposure. " Generator error 60-0313: ILR max current detected". hardware detection of maximum current in serial resonant circuit. It should cause by tube spit, HV tank failed in high load, KV control board defect. " Generator error 60-0316 Spit Max error"kV control has detected the number of tube spits limit from system error logs ( china support Lindewei) " Generator error 60-0310; (3): kV max detected" It means KV reach 160KVP during exposure was detected. The error also cause by KV control board defect or HV tank arcing during exposure. " Generator error 60-0313; (3): ILR max current detected". This is a hardware detection of maximum current in serial resonant circuit. It should cause by tube spit, HV tank failed in high load, KV control board defect. " Generator error 60-0316; (3): Spit Max error". It means kV control has detected the number of tube spits during exposure has reached the limit. It also caus

Applies to

Models
Revolution
Product
Revolution XRd2X
Software
Revolution XR/d 2X Platform (Lightning) – M4
Component
Wallstand (Revolution XR/d 2X) Wallstand (Revolution XR/d 2X) HV TANK, P/N: 2214527 1, HV TANK, P/N: 2214527PROGRAMMED 2, KV CTRL P6.22BF2, P/N:2407577 changed to "5136161-4" JEDI Generator This generator is used on Proteus and Precision 500D systems.
Area
X-ray R&F