Pular para o conteúdo principal
GE Healthcare · CTLightSpeed1848453204718486215006182209182207

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 184845 Cradle unable to move to position 32047, internal cradle error status 1500 see Long Description for full error Fri Sep 1 15:04:42 2006 Error: 184845 Host: ETC Process: Cradle File: cmc interface.c 1.30 9/25/96 Line: 577 Function: No System Function Reported Scan: 1145/2/5 Type: Axial Exception Level: Pri/Most Severe Time: 15:04:41:730 Log Series: 453 Cradle unable to move to position 32047, internal cradle error status 1500

Error: 184862 Cradle unable to initiate a profile. Desired position 3880, velocity 47 internal cradle status 15006 See Long Description for full error Error: 184862 Host: ETC Process: Cradle File: cmc_interface.c 1.30 9/25/96 Line: 715 Function: No System Function Reported Scan: 8637/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 22:05:59:490 Log Series: 104 Cradle unable to initiate a profile. Desired position 3880, velocity 47 internal cradle status 15006

Error: 182209 HW error: No Encoder C pulse within 410 mm (two turns of the Potentiometer). See Long Description for full error Sun Nov 6 12:33:30 2005 Error: 182209 Host: ETC Process: Cradle File: cmcCradlePos.c 1.21 2/5/96 Line: 84 Function: PATIENT ANATOMY POSITIONING : Cradle

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2269175-3 LONGITUDINAL ENCODER ASM (Used on HSA, CTi, H1 QXi, Plus, Ultra, 16 slice H4 and HiSpeed QXi.) replaces 2269175-2. See long description for details Part #: 2269175-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Assembly with Dynapar encoder FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006. ** FE Assist ** If no stock or unable to order the 2269175-3 , order 2269175-2. (212001261 5/22/07) Part #: 2269175-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Asm. NOTE:This asm. has LCC encoder/tested and sensor gap is 0.06-0.08mm FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues