GE Healthcare · CTLightSpeed260003124500422600031255000226000314113155
LightSpeed 4.X - GRE
Symptom
"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors
Error message
Error: 260003124 Aperture Homing Problem See LD for complete error Mon Jan 10 09:04:41 2005 Error: 260003124 Host: HSC Process: CCB Firmware File: aperture_home.c Line: 143 Running diagnostic: Collimator Functional Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50042/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 09:04:41:615 Log Series: 58 Aperture Homing Problem. Error: 260003125 Did not detect a encoder index pulse when homing CAM A. See LD for more details Tue Oct 10 06:52:31 2006 Error: 260003125 Host: HSC Process: CCB Firmware File: aperture_home.c Line: 409 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50002/1/2 Type: Axial Exception Level: Pri/Soft Time: 06:52:31:395 Log Series: 42 Did not detect a encoder index pulse when homing CAM A. Error: 260003141 Filter homing problem. See the Long Description for the full error code Fri Mar 4 21:22:12 2005 Host : Collimator Ermes # : 260003141 Exception Class : Abort Severity : Primary File : filter_home.c Line# : 147 Function : No System Function Reported Scan Type : None/Unknown Scan: 13155/1/0 Filter homing problem. Error: 260003158 The Filter Home Flag Has Moved Since The Last Re
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required