Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT290009015200352908200001821290009030200352906

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

A Pink Pop up message is reported on boot-up that Recon times may be slower - scanning may continue See Long description to see the Pink Pop up message A Pink Pop up message is reported on boot-up that Recon times may be slower

Error message

Error: 290009015 Recon connect timer triggered on 3 ( could also be 1 or 2 ) See LD for full error Error: 290009015 Recon connect timer triggered on 3 Tue Aug 22 07:26:18 2006 Error: 290009015 Host: darc Process: recon_control File: recon_control_cb.cxx Line: 1605 Recon connect timer triggered on X X = the number of ig that failed ( i.e. 1 , 2 or 3)

Error: 200352908 Error occurred while trying to query recon status. See long description for a full sample error. Thu Feb 14 08:28:09 2008 Error: 200352908 Host: oc0 Process: reconSelftest File: reconSelftest.c Line: 623 Error occurred while trying to query recon status. Status = 200001821

Error: 290009030 Failed to stop image_generation process See long description for full error Mon Nov 19 06:31:41 2007 Error: 290009030 Host: darc Process: recon_control File: IGControl.cpp Line: 591 Failed to stop image_generation process.

Error: 200352906 Errors starting up the recon sub system. See Long Description for complete error Mon Oct 10 11:36:18 2005 Error: 200352906 Host: wbvct1 Process: reconSelftest File reconSelftest.c Line: 314 Errors starting up the recon sub system. Status = 1237

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5159834-3 Jarrell VIG with Improved EMC Shielding, replaces 5159834-2 see long description for details 5159834-3 Jarrell VIG with Improved EMC Shielding, replaces 5159834-2 Part #: 5159834-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Jarrell VCT IG with VRAC2, Improved EMC Performance FRU: Yes Status: Orderable Per ECR 2042316, 212008898, 5/8/07. ** FE Assist ** FE must order software release 06MW29.7, p/n 5193989 , via FEMC/Coakly and perform LFC to support this FRU hardware if system is not currently running sw rls 06MW29.7, or more recent ( 5/15/07 212001261)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues