GE Healthcare · CTLightSpeed178517178575212002267244552

LightSpeed 4.X - GRE

Symptom

Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .

Error message

Error: 178517 Failure writing to socket connected to controller ETC. No route to host See LD for complete error Fri Oct 8 07:49:33 2004 178517 4 ct03 fwmgr NBSSocket.c 690 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failure writing to socket connected to controller ETC. OS error : 113 --- No route to host Error: 178517 Failure writing to socket connected to controller ETC. No route to ho

Error: 178575 Failed to R-HARD the following controllers: ETC. See Long Description for complete error. Error: 178575 Failed to R-HARD the following controllers: ETC. Fri Nov 25 13:10:08 2005 Suite: CHCT Host: SBC0 Proc: scse Error: 178575 File: HWinit.m 1.3 Method: resetControllers Line: 384 Failed to R-HARD the following controllers: ETC. Thu Jun 21 08:45:51 2007 Error: 178575 Host: ct01 Process: fwmgr File: HwSetUp.c Line: 837 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to R-HARD the following controllers: STC ETC OBC. 212002267 06/21/2007 13:15 GMT

Error: 244552 ETC controller has failed to reset See the long description for the full error code Wed Aug 24 06:48:26 2005 Error: 244552 Suite: C408 Ho

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-297093P1 AC/DC POWER SWITCH - used on Zeus tables See LD for more info 46-297093P1 HSA-CTI AC/DC POWER SWITCH Part #: 46-297093P1 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: AC/DC POWER SWITCH FRU: Yes Status: Orderable FLS - IBS PART AC/DC TOGGLE SWITCH Comments: FLS - IBS PART AC/DC TOGGLE SWITCH
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required