GE Healthcare · CTLightSpeed29000901329000901629001200429000900931809
LightSpeed 4.X - GRE
Symptom
System fails to reconstruct images. System fails to reconstruct images
Error message
Error: 290009013 Not restarting IG from fd: 31 because we have reached the max number of consecutive resets. See LD for full Error Tue Jul 12 14:31:49 2005 290009013 4 darc recon_control recon_control_cb.cxx 460 Not restarting IG from fd: 31 because we have reached the max number of consecutive resets. Error: 290009016 Recon is going to SHUTDOWN mode due to maximum number of consecutive suspends reached. See long description for complete error Mon Jul 3 10:20:53 2006 Error: 290009016 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 3070 Recon is going to SHUTDOWN mode due to maximum number of consecutive suspends reached. Error: 290012004 RAC hardware failure: error Making image See Long Discription for Details Mon Aug 2 08:37:19 2010 Error: 290012004 Host: ig1 Process: image_generation File: RacInterface.cpp Line: 780 RAC hardware failure: error making image Do not Use - use one of the other Error: 290009009 Image failed due to many reasons See Long Description for complete error Tue Apr 19 12:44:49 2005 Error: 290009009 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 3968 Function: Image Generation : SBC Processing Job Type: RECON_JOB_PROS_AX
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5147443-2 Jarrell IG for NON VCT systems (replaces 5147443 ). See long description for details Part #: 5147443-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Jarrell GRE IG with RAC rev.D or Higher FRU: Yes Status: Orderable Comments: Per ECR 2040301 (Part of Jarrell NPI), 212008898, 3/8/07. **FE Assist** applies to CT but not PET systems. Must order software release 07MW11.10, p/n 5220443 (deployed by FMI 25389), and perform a LFC to support this FRU if system is not currently running sw rls 06MW29.7 or 07MW11.10 (8/16/07 212001261) 212008041 12/03/2007 15:23 GMT Part #: 5147443 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Computer - Image Generation, 2nd Generation FRU: Yes Status: Orderable Comments: Per Jarrell RSPL, 7/19/06, ms11892. ** FE Assist ** If ordering this Jarrell IG to replace a Westville IG 2362872 must also order a 5127222-2 Shielded Network cable and replace the existing one on your system Plus, must order software release 06MW29.7, p/n Part #: 5193989 and perform a LFC. (12/8/05 212001261
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required