GE Healthcare · CTLightSpeedVCT260132796500892601323545014826010102350444
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260132796 The Following Digital Interface Boards Are Missing/Faulty. See Long Description for complete error Mon Feb 6 09:42:29 2006 Error: 260132796 Host: orp Process: orp_subsys File: IfbConfig.cxx Line: 96 Mon Feb 6 09:42:28 2006 Host : Das Ermes # : 260132796 Exception Class : Abort Severity : Secondary File : IfbConfig.cxx Line# : 96 Function : No System Function Reported Scan Type : Static Scan: 50089/1/1 The Following Digital Interface Boards Are Missing/Faulty. Digital Interface Boards: 19 Error 260132354: Scan ABORT -- DAS Has Detected An Error See long description for full error Mon Sep 29 09:01:47 2008 Error: 260132354 Host: orp Process: orp_subsys File: ScanMgr.cxx Line: 611 Mon Sep 29 09:01:46 2008 Host : Das Ermes # : 260132354 Exception Class : Abort Severity : Primary File : ScanMgr.cxx Line# : 611 Function : No System Function Reported Scan Type : Axial Scan: 50148/1/1 Scan ABORT -- DAS Has Detected An Error. Scan Setup Errors Prevent The Data Acquisition System From Being Able To Scan. (See Previous Log Entries.) Do Not Use - use on of the other two Error: 260101023 The (DCB or CCB) has notified the system of an error see long description for examples. Tu
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5955048 (eqvt part 5151006, 5113710) VCT DIFB Digital Interface Board - Sherlock detector See long description for details. Part #: 5955048 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT DIFB PWA FRU: Yes Status: Orderable Parts Offered as: Current: 5955048 Harvested: Not Available Exchange: Not Available Comments: Per GEC-to-Flextronicw transfer program and ECR 2104697, 212008898, 24 Aug 2010. Old FRU Compatible Yes Comments Per GEC to Flextronics Transfer Program #020000, 212008898, 8 Sept 2010. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5955048 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number BOK39368 Old FRU Compatible Yes Comments Per GEC to Flextronics Transfer Program #020000, 212008898, 8 Sept 2010. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5151006 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number BOK39368 5151006 (eqvt part 5113710) VCT DIFB Digital Interface Board - Sherlock detector Part #: 5151006 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT DIFB PWA FRU: Yes Status: Orderable ** FE ASSIST ** If no stock can use P/N 5113710 (212001261 10/12/07) Part #: 5113710 Part Type: GP CONSUMABLES Part Weight: -90719.09 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT DIFB FRU: Yes Status: Orderable ** FE ASSIST ** If no stock can use P/N 5151006 (212001261 10/12/07)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.