GE Healthcare · X-ray R&FRevolution

494F System communication error, retry/reset or call service

Symptom

494F System communication error, retry/reset or call service 494F System communication error, retry/reset or call service.Wallstand Display module no display.

Error message

494F System communication error, retry/reset or call service See long description Error 494F MESSAGES Console Error Message: System communication error, retry/reset or call service Debug Mode Error Message: CANOPEN WS Display Board Failure MODULE Positioner DESCRIPTION The positioner (Master CAN node) cannot communicate with the wall stand display module. RECOMMENDED ACTIONS 1. Check CANOPEN cable connections and terminations. 2. Check wall stand power supply connections and fuses. 3. Check wall stand display connections. 4. Replace wall stand display module. 5. Replace wall stand electronics module.

Applies to

Models
Revolution
Product
Revolution XRd
Software
28.2 Version
Component
Wallstand (Revolution XR/d - Gipeto)
Area
X-ray R&F

Full cause & step-by-step resolution