Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT24452026010100726010101250060260101008260101022

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Attention: The Scanning Hardware is not available. Please see long description. Attention: The Scanning Hardware is not available, please wait for the hardware reset successfull status posted in the feature status area on the right monitor. If the hardware reset fails, then reset the scan hardware from the Service Desktop System Resets button.

Error message

Error: 244520 Scanning hardware has been shut down. NOTE: This is a very generic error. You need to search the log for a more specific error. Fri Apr 21 07:30:53 2006 Error: 244520 Host: oc01 Process: fwmgr File: HwManager.c Line: 1536 Scanning hardware has been shut down. NOTE: This is a very generic error. You need to search the log for a more specific error.

Error: 260101007 CAN serial link error.An CC error was detected when sending a message to the DCB Thu Mar 18 05:36:28 2010 Error: 260101007 Host: orp Process: orp_subsys File: rcib_cc_send.c Line: 373 Thu Mar 18 05:36:28 2010 Host : On-Board Rotating Processor Ermes # : 260101007 Exception Class : Secondary Severity : Sec/Soft File : rcib_cc_send.c Line# : 373 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 CAN serial link error. An CC error was detected when sending a message to the DCB

Error: 260101012 The Rotating Controller Interface Bus (RCIB) fault line has been asserted. See LD for more details Mon Apr 10 07:20:41 2006 Error: 260101012 Host: orp Process: orp subsys Mon Apr 10 07:20:40 2006 Host : On-Board Rotating Processor Ermes # : 260101012 Exception Class : Abort Severity : Primary Fi

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5129366 cable, 48vdc pwr - Fuse board to das, used on vct See long description for details Part #: 5129366 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: cable, 48vdc pwr with special jackscrews, das, vct FRU: Yes Status: Orderable Comments: Added per VCT M4 RSPL Update, ms13452 060206
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues