Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT26011832626011814726011831526011801150075244540

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118326 The communication watchdog with the generator has failed. See long description for a sample error. Thu May 27 07:10:44 2010 Error: 260118326 Host: orp Process: orp subsys File: gen exec.cxx Line: 7521 Exception Class : Abort Severity : Primary Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 The communication watchdog with the generator has failed.

Error: 260118147 (25-0537)LVPS400:main inverter + 26V too high. See long description for a sample error message. Thu May 27 07:10:38 2010 Error: 260118147 Host: orp Process: orp subsys File: gen exec.cxx Line: 2505 Exception Class : Secondary Severity : Pri/Soft Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 (25-0537)LVPS400:main inverter + 26V too high. Z6,Z7= e3h,0h

Error: 260118315 JEDI ID Communication Timer Expired. Check power and commnication interconnects. See Long Description for complete error Mon Dec 6 12:04:49 2004 Error: 260118315 Host: orp Process: orp_subsys File: gen_exec.c Line: 1665 Mon Dec 6 12:05:01 2004 Host : On-Board Rotating Processor Ermes # : 260118315 Exception Class : Soft Severity : Pri/Soft File : gen_exec.c Line# : 1665 Function : No

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
2324118-3 VCT-JH4 Auxiliary Equipped. Aux Box See Long Description for details Part#:2324118-3 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT-JH4 AUXILIARY EQUIPED FRU: Yes Status: Orderable Comments: Set as Critical Part for Product CT_VCT, ms04278, 05Oct28 Added per VCT RSPL, ms13452 041205 Picture below
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues