GE Healthcare · CTLightSpeed200110037
LightSpeed 5.X (80/100kW)
Symptom
Customer gets a forced tube warmup warning pop up. Small Spot will not be used for patient scan Customer gets a forced tube warmup warning pop up. Small Spot will not be used for patient scan. The latest version of SW (07BW08.3.) for the RT and LightSpeed Xtra will give a popup warning and disable the small filament if a tube warmup needs to be completed.
Error message
Error: 200110037 The Forced Tube Warmup Warning Popup, Small Spot will not be used for patient scan See Long Description for entire Error Mon Jul 9 09:07:07 2007 Error: 200110037 Host: prd_ge_ct04 Process: scanRx File: ZeusrpGlobalActions.c Line: 2223 The Forced Tube Warmup Warning Popup, Small Spot will not be used for patient scan .
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT