GE Healthcare · CT
Sytec 1600i/1800i SYj Eu
Symptom
Unable to scan with XG errors
Error message
ER-102 Gantry Error code 4 see LD ER–102 Gantry Error code 4 0000000X Error Code X = 1: Gantry Request Command Error 2: Gantry Set Timeout in 20 sec 3: Servo Motor Overheat 4: Servo Amp Alarm 5: Gantry is not in System Mode 6: Gantry Rotation Not Response 7: Gantry Rotation Overspeed 8: Gantry Rotation Underspeed 9: Safety Switch Open A: Dynamic Brake ON B: Gantry Initialize Error or Position Error F: GNTINITREQ Timeout For ER-102 code 4 Servo Amplifier showed seven segment LED code 1 for Over current.
Applies to
- Product
- Sytec 1600i/1800i SYj Eu Sytec 1600i/1800i SYj Eu Sytec 1600i/1800i SYj Eu Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2112365 Servo Amplifier used on Sytec 1800i See long description Part #: 2112365 Part Type: GP CONSUMABLES Part Weight: -499995.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: P9213SB SERVO AMP. FRU: Yes Status: Orderable Comments: FLS - SERVO AMP USED ON CT SYJ GANTRY. NOT REPAIRABLE PER YMS. PRODUCT SOLD IN LATIN AMERICA BUT NOT THE U.S. STA 8/21/95
- Area
- CT