Symptom
Fastcal fails the mylar window check. Fastcal fails the mylar window check
Error message
Error: 230023033 Mylar Window Check has failed. Mylar Window may be dirty. Calibration vectors generated with the dirty mylar window may cause Image artifacts. See Long Description for complete error Sat Jan 8 08:06:05 2005 230023033 Host: uhct Process: dailyPrepRx File: DailyPrepUtils.c Line: 2967 Mylar Window Check has failed. Mylar Window may be dirty. Calibration vectors generated with the dirty mylar window may cause Image artifacts. Error: 230017088 The mylar window check failed See Long Description for complete error Mon Aug 23 05:50:31 2004 Error: 230017088 Host: usafaultra Process: accs File: aircal processing.c Line: 1025 The mylar window check failed - Scan 50019/1/3
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5142849-4 SCAN WINDOW LIGHTSPEED SERIES H2 THRU HP RIM COVERS (replaces 2101967-4). See long description for details. Part #: 5142849-4 Part Type: GP CONSUMABLES Part Weight: 7.94 LBR Repairable: No Returnable: No Hazardous Material: N Description: ASSEMBLY, SCAN WINDOW, LIGHTSPEED SERIES H2 THRU HP RIM COVERS, WEATHERSTRIP UPDATED MAY 2005 FRU: Yes Status: Orderable Comments: This part is replacement for 2101967-4, please discontinue use of 2101967-4 and use only 5142849-4. Added per ECR2014906, ms13452 050713
- Area
- CT
Full cause & step-by-step resolution