Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT23000090053852633554432123321812672703175000

LightSpeed 3.X (Ultra, Plus, QX/i) - GRE

Symptom

"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors

Error message

Error: 230000900 WARNING slow write to disk See llong decsriptions for full error Thu Jan 29 08:54:34 2009 Error: 230000900 Host: darc Process: dataacq File: file_save_thread.cxx Line: 194 WARNING slow write to disk: 51.538526 sec to write 33554432 bytes; scan CT07/152/3/3-1233218126/727031; scan seq 75000 Example: This took 51 seconds to write 33 mb. which is extremely long. 51.538526 sec to write 33554432 bytes;

Error: 230000897 Write to scan data disk failed. See LD for more details Tue May 23 07:22:28 2006 Error: 230000897 Host: darc Process: dataacq File: raw_file.cxx Line: 250 Function: Data Acquisition : Scan Data Save Scan Type: Not defined Scan: 50074/1/1 Scan Seq Id: 0 Write to scan data disk failed. Status: -1 Most Recent System Error: OS error : 98 --- Address already in use

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5144058-501 Upgrade Kit - SDDA and GOC3 - this part is no longer orderable. See SN11063012 - Order 5144058-500 plus the proper DARC2 See Long description 5144058-501 Upgrade Kit - SDDA and GOC3 MDAS 4,8 with DARC2 Replacement Kit, JARRELL Part #: 5144058-501 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Service Kit - SDDA and GOC3 4,8 MDAS DARC Replacement Kit, JARRELL FRU: Yes Status: Orderable This kit upgrades a GOC3 console to a GOC4 console with DARC 2 and eliminates the SDDA . FE ASSIST ** Currently for CT not for Pet Must also order 5193989 and perform a LFC to support this FRU if system is not currently running 06MW29.7 or better.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.