GE Healthcare · CTHiSpeed186507
HiSpeed Dual, ProSpeed FII Series
Symptom
Unable to scan with "collimator request time out error". see long description. Unable to scan with "collimator request time out error".
Error message
Error: 186507 Z-Axis Collimator Error. Code : 30-1111-07 The number of pulses for Z-Axis Motor is under specification when moving from FWD limit to BWD limit on Home Positioning. See LD for complete error Mon May 9 18:03:19 2005 Error: 186507 Host: CTE Process: tgp_out File: tgp_in_input.c Line: 449 Z-Axis Collimator Error. Code : 30-1111-07 Error detected by : CIF The number of pulses for Z-Axis Motor is under specification when moving from FWD limit to BWD limit on Home Positioning.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed Dual, ProSpeed FII Series HiSpeed Dual, ProSpeed FII Series Product Name: Hispeed Dual ( Linux ) CT/Be2 A list of types. (a Product/hardware table) This console has an HP8000, HP8200, or HP8400 Computer (not GRE console) An On-site / Remote identification aid . (Example: check_config) HP Host computer A Table of options (included, optional, new introductions, …) HiSpeed Dual for China and Global B71402CA has the following as standard Options 1.5 sec fast scan, Power 200ma, Thichness 1mm, Raw data 5000, Image 100000 A list of Software that has run on this product (Product / Software Compatibility Table) With HP8000/8200 : • OS Version 4.5.15144980•Software Version 2.235165193 (except China)•Software Version 2.235172510 (for China only)•Patch 2.245179238•Patch 2.255197828•Patch 2.275267772 With HP8400 : •OS Version 6.2.95265475•Software Version 3.105270457•Patch 3.305267336•DVD Interchange 6.15270187•Patch 3.40 (See Note 8)5316222•DVD Interchange 7.05316345N/ Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links: Drawing Print 2320306-2CCT : http://gemspdm.med.ge.com/ematrix/common/eLibDocument.jsp?type=type_DrawingPrint&name=2320306-2CCT&rev=9 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2258540 P2014NJ (also 2258540-2) OPTO ASSY - Used on YMS HiSpeed Collimator See long description for details Part #: 2258540 Part Type: GP CONSUMABLES Part Weight: 0.04 KGM Repairable: No Returnable: No Hazardous Material: N Description: P2014NJ SNSR ASSY FRU: Yes Status: Orderable ====================== Parts Offered as: Current: 2258540 Harvested: Not Available Exchange: Not Available Comments: ** FE ASSIST ** For CT Hispeed Dual For GDAS Gantry, 2258540-2 can replace 2258540. 2258540-2 is backward compatible with 2258540. For MDAS Gantry, 2258540-2 and 5120153-2 should be used together to replace 2258540 (212001261 7/13/08) FLS - Z AXIS SENSOR ASM U/O EBISU (NXI) GANRY MS08992 6/5/00
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.