Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed26011818426000740250444260118119

BrightSpeed Select, Delight Select or MERI Select

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118184 (50-0203) Heater inverter permanent overcurrent.(SW limit) See LD for error explanation Error: 260118184 (50-0203) Heater inverter permanent overcurrent.(SW limit) Code 50 Filament Heater Errors Error code: 50-0203H Message/explanation: Heater inverter permanent over current.(SW limit) Issued by the heater board when an over current has been detected and 3 restarts have been tried without success within 100 ms Potential cause: Heater board Recommended action/ Troubleshooting guide: Replace Heater board

Error: 260007402 Converter Board(s) Asserted a Board FAULT. See Long Description for complete error Tue Dec 28 09:41:31 2004 Error: 260007402 Host: orp Process: orp_subsys File: cnvmgr.c Line: 605 Tue Dec 28 09:41:30 2004 Host : Das Ermes # : 260007402 Exception Class : Abort Severity : Secondary File : cnvmgr.c Line# : 605 Function : No System Function Reported Scan Type : Static Scan: 50444/1/1 Converter Board(s) Asserted a Board FAULT.

Error: 260118119 (20-0254)Heater Inverter overcurrent(inverter1) See Long Description for full error Sat Apr 15 11:57:29 2006 260118119 orp orp subsys gen exec.c 2150 Sat Apr 15 11:57:27 2006 Host : Jedi Ermes _ : 260118119 Except

Applies to

Models
BrightSpeed
Product
BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2361419 Programmed Heater Board (eqvt 2409049) inside Jedi Auxiliary Module 2324118 See LD for Part Details - please read the note at the bottom. Part number for Heater (Filament) Board below: Part #: 2361419 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED HEATER 6P04 FRU: Yes Status: Orderable Part #: 2409049 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: FRU-PROGRAMMED HEATER V6.04 FRU: Yes Status: Orderable Parts Offered as: Current: 2409049 Harvested: Not Available Exchange: Not Available Comments: added per to ECR2088514 by sso#305010201 in Jan 15, 2010 Note: * Order 2409049 to replace 2361419.
Area
CT

Full cause & step-by-step resolution