LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

A pink Attention box pops up during Fast Cal saying there is a problem with the DAS/Detector. Scanning will NOT be disabled. Image Quality will be compromised, call GE SERVICE. See long description for pop up message picture. A pink Attention box pops up during Fast Cal saying there is a problem with the DAS/Detector. Scanning will NOT be disabled. Image Quality will be compromised, call GE SERVICE. After clicking OK Fast Cal completes. ATTENTION: The system detected a problem with the DAS/Detector sub-system. Image Quality will be compromised. Scanning will NOT be disabled, but it is critical that the user call GE SERVICE to rectify the problem. Press 'OK" to proceed at the risk of Image Artifacts

Error message

Error: 230007004 Channel ###, Module ## Short to ground* FAILED X- AXIS EXCLUSION CRITERIA EXCLUSION CRITERIA *. Error: 230007004 can be used for any row, channel and module that is failing. Error: 230007004 Channel ###, Module ## Short to ground* FAILED X- AXIS EXCLUSION CRITERIA Examples: Sat Sep 17 08:08:02 2005 Error: 230007004 Host: amhvct Process: badchanmgr File: BadChannelMap.cxx Line: 584 Exclusion Test - Row 33:1B, Channel 900, Module 57 Short to ground* FAILED X-AXIS EXCLUSION CRITERIA *. Tue Aug 29 05:24:31 2006 Error: 230007004 Host: ct99 Process: badchanmgr File: BadChannelMap.cxx Line: 667 Exclusion Test - Row 1:32A, Channel 418, Module 27 Short to ground* FAILED X-AXIS EXCLUSION CRITERIA *.

Error: 230007007 Exclusion Test - Row 33:1B, Channel 899, Module 57 Short together* FAILED PROXIMITY CRITERIA *. Sat Sep 17 08:08:02 2005 Error: 230007007 Host: amhvct Process: badchanmgr File: BadChannelMap.cxx Line: 627 Exclusion Test - Row 33:1B, Channel 899, Module 57 Short together* FAILED PR OXIMITY CRITERIA *.

Error: 230007008 Exclusion Test - Row 33:1B, Channel 912, Module 57 Short to ground* FAILED PROXIMITY CRITERIA * Sat Sep 17 08:08:03 2005 Error: 230007008 Host: am

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5955048 (eqvt part 5151006, 5113710) VCT DIFB Digital Interface Board - Sherlock detector See long description for details. Part #: 5955048 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT DIFB PWA FRU: Yes Status: Orderable Parts Offered as: Current: 5955048 Harvested: Not Available Exchange: Not Available Comments: Per GEC-to-Flextronicw transfer program and ECR 2104697, 212008898, 24 Aug 2010. Old FRU Compatible Yes Comments Per GEC to Flextronics Transfer Program #020000, 212008898, 8 Sept 2010. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5955048 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number BOK39368 Old FRU Compatible Yes Comments Per GEC to Flextronics Transfer Program #020000, 212008898, 8 Sept 2010. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5151006 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number BOK39368 5151006 (eqvt part 5113710) VCT DIFB Digital Interface Board - Sherlock detector Part #: 5151006 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT DIFB PWA FRU: Yes Status: Orderable ** FE ASSIST ** If no stock can use P/N 5113710 (212001261 10/12/07) Part #: 5113710 Part Type: GP CONSUMABLES Part Weight: -90719.09 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT DIFB FRU: Yes Status: Orderable ** FE ASSIST ** If no stock can use P/N 5151006 (212001261 10/12/07)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.