Symptom

The table is frozen, will not move THE TABLE IS FROZEN/WILL NOT MOVE

Error message

Error: 260125044 CC receive failed. status:112 MsgCode:0. NodeId:21 msgSize:9010162 bufSize:632.Probable software problem. Error: 260125044 CC receive failed. Probable software problem. Sun Aug 8 19:55:55 2010 Error: 260125044 Host: tgp Process: tgp_subsys File: ta_main.c Line: 397 Sun Aug 8 19:55:54 2010 Host : Table Gantry Processor Ermes # : 260125044 Exception Class : Secondary Severity : Sec/Soft File : ta_main.c Line# : 397 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 CC receive failed. status:112 MsgCode:0. NodeId:21 msgSize:9010162 bufSize:632. Probable software problem.

Error: 244633 Table reports that Table Communication becomes down. Wed May 25 17:35:00 2005 Error: 244633 Host: ct01 Process: fwmgr File: fwmgr_events.c Line: 7301 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Table reports that Table Communication becomes down.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.