GE Healthcare · CTLightSpeed26000793626000777426000740250444
LightSpeed 5.X (60kW)
Symptom
Scan abort during alpha calibration. Scan abort during alpha calibration
Error message
Error: 260007936 DAS Converter Board(s) encountered an error during alpha calibration. See LD for complete error which shows failing board number Error: 260007936 Host: orp Process: orp_subsys File: \ Host : Das Ermes # : 260007936 Exception Class : Soft Severity : Pri/Soft File : brdflt.c Line# : 82 Function : No System Function Reported Scan Type : None/Unknown Scan: DAS Converter Board(s) encountered an error during alpha calibration. See Following Error Message for Converter Card List Board :6 ASIC/Channel :0x0137 Stage :1 Board Channel :19 Error: 260007774 List of Converter Cards From the Previous Error Message Card List: xx xx See Long Description for full error Tue Dec 30 09:25:17 2008 Error: 260007774 Host: orp Process: orp_subsys File: CnvBoardInterface.cc Line: 852 Tue Dec 30 09:25:17 2008 Host : Das Ermes # : 260007774 Exception Class : Secondary Severity : Sec/Soft File : CnvBoardInterface.cc Line# : 852 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 List of Converter Cards From the Previous Error Message Card List: 8,11,14-15,19,21,25-27,29-36,38-58,60-63 Note on this error. Sometimes the actual bad board will be the one that is not listed
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required