Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed2601152452375517710

BrightSpeed HPower, Delight, Velocity or MERI HPower

Symptom

Cradle will not drive in or out. Cradle will not drive in or out

Error message

Error: 260115245 MTCB cradle encoder delta data out of range. See long description for full sample error. Thu Aug 7 08:34:27 2008 Error: 260115245 Host: tgp Process: tgp subsys File: mtcb cradle monitor.c Line: 285 Thu Aug 7 08:34:27 2008 Host : Table Sub-System Ermes # : 260115245 Exception Class : Secondary Severity : Pri/Soft File : mtcb cradle monitor.c Line# : 285 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 MTCB cradle encoder delta data out of range. Possible Encoder, cabling or MTCB issue. cradle current position = 23755 cradle previous position = 17710 Pacsci drive command position = -96 position offset = -396 previous position offset = -396

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2339120 Cradle drive assembly See long decription for details Part #: 2339120 Part Type: GP REPAIRABLES Part Weight: 2.20 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: CRADLE DRIVE ASSY - H PWR FRU: Yes Status: Orderable Comments: FLS - CRADLE DRIVE ASSY - H PWR. ADDED PER HPOWER PROGRAM RSPL BY MS20544 ON 2/24/03. POSSIBLE M3 DATE: FW19, 2003.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues