GE Healthcare · CTHiSpeedLightSpeed26000745750084
"Scan hardware Stopped Scan" with beam tracking errors.
Symptom
"Scan hardware Stopped Scan" with beam tracking errors. "Scan hardware Stopped Scan" with beam tracking errors
Error message
Error: 260007457 Beam Tracking Control Loop Error Exceeded The ABORT Tolerance See Long Description for complete error Wed Apr 27 18:54:40 2005 Error: 260007457 6 DAS DCB Firmware zaxis.c 564 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50084/6/0 Type: None/Unknown Exception Level: Pri/Soft Time: 18:54:40:490 Log Series: 936 Beam Tracking Control Loop Error Exceeded The ABORT Tolerance. Tolerance : 400 umd Control Error : 448 umd Beam Side : B Side
Applies to
- Models
- HiSpeed, LightSpeed
- Product
- HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required