Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT210000447210000428210000431210000430200002357

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Applications software fails to boot up completely and automatically shuts itself back down. Applications software fails to boot up completely and automatically shuts itself back down

Error message

Error: 210000447 unable to connect to session dms_session with error code -2 See long description for details Tue Jan 29 06:51:29 2008 Error: 210000447 Host: gevct2 Process: dms File: DMFInstall.cxx Line: 229 Error: unable to connect to session dms_session with error code -2 error message Session Not Connected

Error: 210000428 Unable to establish connection for repository iosterra|terradb session See long description for full error Tue Jan 29 06:51:29 2008 Error: 210000428 Host: gevct2 Process: dms File: Initialize.cxx Line: 396 Unable to establish connection for repository iosterra|terradb session

Error: 210000431 Failed to intialize client session for save context with error 2 See long description for sample error. Tue Mar 18 08:58:45 2008 Error: 210000431 Host: jghls64 Process: dms File: SaveContext.cxx Line: 41 Failed to intialize clien session for save context with error 2

Error: 210000430 Failed to initialize SaveContext with status -1 See long description for sample error. Tue Mar 18 08:58:45 2008 Error: 210000430 Host: jghls64 Process: dms File: Initialize.cxx Line: 410 Failed to initialize SaveContext with status -1

Error: 200002357 A HOST SHUTDOWN was initiatied by th

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Software
08MW33.2 for GOC6 LightSpeed VCT (64 Slice Only) 11/17/2008 Inital M3 release. This software is now obsolete, this site should be upgraded to 11HW12.5 software released via FMI 25427 08MW33.2 for GOC6 LightSpeed VCT (64 Slice Only) 11/17/2008 Inital M3 release. LightSpeed 7.1 ( 08MW33.2 ) Software Release supports GOC6 consoles used on VCT 64, VCT XT & VCT Std. - LightSpeed 7.1 (08MW33.2)Software for GOC6 Collector 5318369 Rev 1 Includes OS and Application CD's - Operating System Linux 5.2.13 Software CD 5318646 Rev 1 - Application LightSpeed VCT GOC6 DVD 5317935 Rev 1 - Class C LightSpeed 7.1 5318761 Rev 1 - Class M LightSpeed 7.1 5317906 Rev 1 An On-site/Remote identification aid . (Example: check_config) Result from check_config or swhwinfo 08MW33.2 V40_PS_HD64_G_GTM 08MW33.2 V40_PS_HD64_G_GTL Please create an ITRAK report for any software related issues
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.