LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260003149 Illegal distance for CAM A encoder count conversion. Could be Cam B also See LD for complete error codes Fri Oct 12 12:43:36 2001 Error: 260003149 Host: HSC Process: CCB Firmware File: cfg_cam.c Line: 191 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50267/1/2 Type: Static Exception Level: Sec/Soft Time: 12:43:36:650 Log Series: 57 Illegal distance for CAM A encoder count conversion. Requested distance: 15799 micrometers Mon Oct 2 10:06:25 2006 Error: 260003149 Host: HSC Process: CCB Firmware File: cfg_cam.c Line: 191 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50646/4/40 Type: Axial Exception Level: Sec/Soft Time: 10:06:25:150 Log Series: 969 Illegal distance for CAM B encoder count conversion. Requested distance: -259 micrometers

Do Not Use - use one of the other two Error: 260001023 See Long Description for complete error NOTE: The error message code included in this error is variable. You will need to refer to the log to find the cause of this error. Tue Dec 7 07:23:22 2004 Error: 260001023 Host: OBC Process: HOP File: hop_erlog.c Line: 367 Function: No System Function Reported Scan: 50723/1/1 Type: Axial Exception Lev

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.