GE Healthcare · CTLightSpeed26012504113078281308322429496729526011519171564
LightSpeed 5.X (80/100kW)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260125041 Motion without command in Cradle drive See Long Description for an example error message. Tue Aug 21 12:19:15 2007 Error: 260125041 Host: tgp Process: tgp subsys File: ta callback.c Line: 507 Tue Aug 21 12:19:14 2007 Host : Table Gantry Processor Ermes # : 260125041 Exception Class : Abort Severity : Primary File : ta callback.c Line# : 507 Function : No System Function Reported Scan Type : None/Unknown Scan: 4747/3/0 Motion without command in Cradle drive Current pos = 1307828 Last stop pos = 1308322 Tolerance = 250 Time since last stop = 20 Last target pos = 4294967295 Hardware status = 8104 Error: 260115191 Cradle position is out of tolerance See long description for full error Thu Sep 1 09:04:12 2005 Error: 260115191 Host: tgp Process: tgp_subsys File: mtcb_cradle_monitor.c Line: 414 Thu Sep 1 09:04:12 2005 Host : Table Sub-System Ermes # : 260115191 Exception Class : Soft Severity : Pri/Soft File : mtcb_cradle_monitor.c Line# : 414 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Cradle position is out of tolerance. cradle_position(um) : 1666 target_position(um) : 71564 velocity (um/s) : 29999 fpga status : 0x8140
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2339120 Cradle drive assembly See long decription for details Part #: 2339120 Part Type: GP REPAIRABLES Part Weight: 2.20 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: CRADLE DRIVE ASSY - H PWR FRU: Yes Status: Orderable Comments: FLS - CRADLE DRIVE ASSY - H PWR. ADDED PER HPOWER PROGRAM RSPL BY MS20544 ON 2/24/03. POSSIBLE M3 DATE: FW19, 2003.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.