Pular para o conteúdo principal
GE Healthcare · CTDiscovery1111710008

Technologist CANNOT remove Exams from the system browser.

Symptom

Technologist CANNOT remove Exams from the system browser. Unable to remove images until Reboot is completed

Error message

Error: 11117 Remote does not support push for SOP_Class_UID See Long Description Two version of this error Exist. “DICOM Enhanced SR SOP” class. UID 1.2.840.10008.5.1.4.1.1 .88.22 “DICOM X-ray Radiation Dose SR SOP” class UID 1.2.840.10008.5.1.4.1.1 . 88.67 Fri Feb 13 15:22:35 2009 Error: 11117 Host: msec8x1 Process: dcserver File: dcspush.c Line: 2456 Remote does not support push for SOP_Class_UID 1.2.840.10008.5.1.4.1.1.88.67 , so all images with this SOP_class_UID have been rejected Fri Feb 13 15:22:35 2009 Error: 11117 Host: msec8x1 Process: dcserver File: dcspush.c Line: 2456 Remote does not support push for SOP_Class_UID 1.2.840.10008.5.1.4.1.1.88.22 , so all images with this SOP_class_UID have been rejected “DICOM Enhanced SR SOP” class. UID 1.2.840.10008.5.1.4.1.1 .88.22 “DICOM X-ray Radiation Dose SR SOP” class UID 1.2.840.10008.5.1.4.1.1 . 88.67 * SR stand for "Structured Report"

Applies to

Models
Discovery
Product
Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
09MW29.14 for Discovery CT750 HD 09MW29.14 for Discovery CT750 HD FMI 25416 09MW29.14.V40_PS_HD64_G_GT - Discovery CT750 HD Collector 09MW29.14 5344429 Rev 3 includes OS and Application CD's - Operating System Linux 5.3.10 Software CD 5344428 Rev 1 - Application CT750 HD DVD 09MW29.14 5344425 Rev 3 ======================================================= - Class C CT750 5344426 Rev 3 - Class M CT750 5344427 Rev 3 An On-site / Remote identification aid . (Example: check_config) Result from check_config or swhwinfo: 09MW29.14.V40_PS_HD64_G_GTM 09MW29.14.V40_PS_HD64_G_GTL
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues