GE Healthcare · CTLightSpeed26000312450042260003099260003120
LightSpeed 5.X (80/100kW)
Symptom
"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors
Error message
Error: 260003124 Aperture Homing Problem See LD for complete error Mon Jan 10 09:04:41 2005 Error: 260003124 Host: HSC Process: CCB Firmware File: aperture_home.c Line: 143 Running diagnostic: Collimator Functional Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 50042/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 09:04:41:615 Log Series: 58 Aperture Homing Problem. Error : 260003099 CAM A Motion With Out Command Exceeded The Warning Tolerance. See Long Description for details. Error : 260003099 CAM A Motion With Out Command Exceeded The Warning Tolerance. Sat May 9 01:20:54 2009 Error: 260003099 Host: orp Process: orp_subsys File: ccb_drive_timers.c Line: 296 Sat May 9 01:20:53 2009 Host : Collimator Ermes # : 260003099 Exception Class : Secondary Severity : Pri/Soft File : ccb_drive_timers.c Line# : 296 Function : No System Function Reported Scan Type : Axial Scan: 8676/2/9 CAM A Motion With Out Command Exceeded The Warning Tolerance. Desired Position : 1599um 562cnt Current Position : 1575um 565cnt Warning Tolerance: 22um Error: 260003120 An error occurred attempting to home CAM A or B See LD for complete error Mon Jan 10 09:04:41 2005 Error: 26000312
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-296259P6 Rotating Coupling See Long Description for FEMC and pics 46-296259P6 ROCOM COUPLING_BORES 5 MM pictures Part #: 46-296259P6 Part Type: GP CONSUMABLES Part Weight: 0.10 LBR Repairable: No Returnable: No Hazardous Material: N Description: ROCOM COUPLING_BORES 5 MM, 0.250 IN. FRU: Yes Status: Orderable Comments: FLS - ADAPTER ADDED PER PCN #220365.USED ON GANTRY. PER PCN COMMENTS, FRU WILL BE CHANGED FROM N TO 1. MSA8429 01/15/02 COUPLING OF 2 SHAFTS FOR COLLIMATOR MOTOR MS10147 01/02 FOR ROTATING APTERETURE
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required