Symptom

Losing communications with ORP when scanning at 0.4 seconds scan speed

Error message

Error: 260100149 Early ORP Exposure Command OFF signal based on ScanControl Scanning Status. Tue Mar 11 09:20:25 2008 Error: 260100149 Host: tgp Process: tgp_subsys File: scGantryInMail.c Line: 493 Tue Mar 11 09:20:25 2008 Host : Table Gantry Processor Ermes # : 260100149 Exception Class : Abort Severity : Primary File : scGantryInMail.c Line# : 493 Function : No System Function Reported Scan Type : Axial Scan: 9341/2/2 Early ORP Exposure Command OFF signal based on ScanControl Scanning Status. Rx Scan Duration:237 Hdw Scan Duration:129 (msec). Trigger Freq:2460 Axial:350. SC: ExpTime:130 (msec). ORP: ExpTime:125 msec

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5120819-2 RoHS Compliant Slipring Brushblock Assembly used on 5.X-7.X gantry. replaces 2349598 5120819 See long description for details 5120819-2 RoHS Compliant Slipring Brushblock Assembly Old FRU Compatible Yes Comments Per ECR 2087298, 212008898, 8/5/09. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5120819-2 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number ECR2087298 Part #: 5120819 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT Slipring Brushblock. Same as Hpower, but has additional voltage creepage distance for 208VAC. FRU: Yes Status: Orderable Comments: Added per ECR 2011062 new VCT Brushblock (5120819) on sliprings 2320265 and 2364819. Also added by VCT NPI RSPL UPDATE, ms13452 050916.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.