GE Healthcare · CTHiSpeed23001708850019
Fastcal fails the mylar window check.
Symptom
Fastcal fails the mylar window check. Fastcal fails the mylar window check
Error message
Error: 230017088 The mylar window check failed See Long Description for complete error Mon Aug 23 05:50:31 2004 Error: 230017088 Host: usafaultra Process: accs File: aircal processing.c Line: 1025 The mylar window check failed - Scan 50019/1/3
Applies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5142849-4 SCAN WINDOW LIGHTSPEED SERIES H2 THRU HP RIM COVERS (replaces 2101967-4). See long description for details. Part #: 5142849-4 Part Type: GP CONSUMABLES Part Weight: 7.94 LBR Repairable: No Returnable: No Hazardous Material: N Description: ASSEMBLY, SCAN WINDOW, LIGHTSPEED SERIES H2 THRU HP RIM COVERS, WEATHERSTRIP UPDATED MAY 2005 FRU: Yes Status: Orderable Comments: This part is replacement for 2101967-4, please discontinue use of 2101967-4 and use only 5142849-4. Added per ECR2014906, ms13452 050713
- Area
- CT
Full cause & step-by-step resolution