GE Healthcare · CTLightSpeedVCT2900090021183997163967569200270032200110026

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

System takes a scout but the tech cannot see it in ExamRx - document in RFS - critical issue System takes a scout but the tech cannot see it in ExamRx.

Error message

Error: 290009002 Error failed in opening file See long description for full error Mon Jul 9 17:34:45 2007 Error: 290009002 Host: darc Process: recon_control File: IsrUpdateCache.cxx Line: 153 Error failed in opening file. file name = /usr/g/queue/priority3_jobs/FCCC.3776.2.1177.1183997163.967569.pro_cine error msg = No such file or directory

Error: 200270032 GraphicRx Error Trace See long description for details Tue Sep 23 08:15:25 2008 Error: 200270032 Host: vct53 Process: graphicRx File: /bat_pool/work3072/commondisplay/src/cxx/ Line: 1188 GraphicRx Error Trace

Error: 200110026 OC Processing Can not find scout scans See long description for full error Sun Dec 23 16:25:29 2007 Error: 200110026 Host: vct1 Process: scanRx File: ViewEdit.cxx Line: 9807 Function: Data Acquisition : OC Processing Can not find scout scans. AutoMa will be disabled for Series:3

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.