GE Healthcare · CTLightSpeedVCT230017078503421111522307108580230017083385870
LightSpeed 1.X - GRE
Symptom
Image artifact, intermittent multiple partial rings See long description for details Image artifact, intermittent multiple partial rings The artifact is easy to see when using Lung window and levels The 4 sequential (1,2,3,4) images below illustrate how on helical scans the partial ring moves around the patient. Image 1 Image 2 Image 3 Image 4
Error message
Error: 230017078 Aircal failed to process scan See Long Description for complete error Tue Mar 22 20:12:04 2005 Error: 230017078 Host: ct01 Process: accs File: accs_events.cxx Line: 463 Aircal failed to process scan - Exam = 50342, Series = 1, Scan = 1, Seconds = 1111522307 Microseconds = 108580 View scanfileto check data and header information Error: 230017083 Possible beam obstructed - sin/cos vector magnitude = 0.xxxxxx - may include row and channel numbers See LD for complete error Mon Oct 31 07:08:03 2005 Error: 230017083 Host: cpcvct2p Process: accs File: aircal_spectral_acal_util.c Line: 445 Possible beam obstructed - sin/cos vector magnitude = 0.385870 Mon Mar 21 07:27:12 2011 Error: 230017083 Host: sj01_ct64s_6001 Process: accs File: aircal_spectral_acal_util.c Line: 449 Possible beam obstructed - row 58(26B), detector channel 375, sin/cos vector magnitude = 0.381015
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2218158 SDAS +/-5VDC ANALOG PWR SUPPLY W/BRACKET. used on 1.x with H1 gantry See Long Description for details Part #: 2218158 Part Type: GP CONSUMABLES Part Weight: 1.81 LBR Repairable: No Returnable: No Hazardous Material: N Description: DAS 5VDC ANALOG PWR SUPPLY W/BRACKET FRU: Yes Status: Orderable Comments: FLS - DAS +/-5VDC ANALOG PWR SUPPLY W/BRACKET U/O HELIOS DAS REPLACES 46-251198P46 MS08992 10/5/98 GENCO INDUSTRIES-WAUKESHA,WI PER FE - 46-251198P46 IS NOT ENTIRELY COMPATIBLE - 3 CONNECTION POINTS FOR WIRES - NEED 4- CANNOT SOLDER WIRE IN CORRECTLY DUE TO HOLE IN SOLDER POINT BEING TOO SMALL FOR MORE THAN 1 WIRE. TW/01/03
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.