GE Healthcare · MR2225518
2225518 Dynamic disable and T/R driver fault due to defective coil.
Symptom
Cannot Scan with Head T/R Driver Error. The system failed to scan either head or body scans. The SSM light indicated T/R Fault.
Error message
2225518 Dynamic disable and T/R driver fault. Possibly a shorted diode in the head T/R switch network, or a short circuit load. See long description Thu Feb 6 04:02:38 1997 2225518 1 lx-bay14b NSP HPC_PORT_ID:spi_tr_mon 6060 Dynamic disable and T/R driver fault. Possibly a shorted diode in the head T/R switch network, or a short circuit load.
Applies to
- Product
- HF OpenSpeed HFO1 HFO2 and HFO3 Only.
- Software
- HFO3
- Component
- 2396079-2 0.7T NV ARRAY COIL WITH CABLE
- Area
- MR
Full cause & step-by-step resolution