Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2600073915010232719

LightSpeed RT16 & RT16 Xtra

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260007391 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply. Same error for -5v Analog DAS Power Supply. Error: 260007391 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply. Same error # for -5v Analog DAS Power Supply. Error: 260007391 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply Wed Feb 22 08:50:01 2006 Host : Das Ermes # : 260007391 Exception Class : Abort Severity : Secondary File : pwrtst.c Line# : 90 Function : No System Function Reported Scan Type : None/Unknown Scan: 50102/3/0 The Data Acquisition System Has Detected An Error In The +5V Analog DAS Power Supply . Measured: 1590 mV, Expected range: 4750 to 5250 mV Tue Aug 22 15:51:09 2006 260007391 4 orp orp_subsys pwrtst.c 90 Tue Aug 22 15:51:08 2006 Host : Das Ermes # : 260007391 Exception Class : Abort Severity : Secondary File : pwrtst.c Line# : 90 Function : No System Function Reported Scan Type : None/Unknown Scan: 32719/3/0 The Data Acquisition System Has Detected An Error In The -5V Analog DAS Power Supply . Measured: -4748 mV, Expected range: -4751 to -5249 mV

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2287134 Mdas or 2287134-4 GDAS Jumper Cable Right or Left, connects to Center Chassis See Long Description for IMPORTANT differences between these two part numbers ! ! 2287134 MDAS or 2287134-4 GDAS Jumper Cable Right or Left side of GDAS that interconnects with the Center chassis NOTE: Check your DAS part number before ordering! 2287134 THIS PART IS FOR MDAS 2358139 ONLY 2287134-4 THIS PART IS FOR GDAS 2358139-2 ONLY Part #: 2287134-4 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Harness, GDAS Jumper, ED.2 EMC FRU: Yes Status: Orderable Comments: PART ADDED PER VCT RSPL AND IS EMC2 COMPLETE 12/01/2004 Part #: 2287134 Part Type: GP CONSUMABLES Part Weight: 0.66 LBR Repairable: No Returnable: No Hazardous Material: N Description: MDAS JUMPER HARNESS FRU: Yes Status: Orderable Comments: FLS - MDAS JUMPER HARNESS - RT/LT ADDED VIA PCN 220296. VENDOR CHANGED FROM NORTECH TO SANMINA SCI IN MEXICO PER E-NPI TRANSFER "SANMINA CABLE TRANSFER". MS20544 ON 6/7/02.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.