System will not scan reporting CIF Zero DAS data was detected.
Symptom
System will not scan reporting CIF Zero DAS data was detected. See Long Description for Details. System will not scan reporting CIF Zero DAS data was detected. Error: 30-1010-01
Error message
Error: 186518 code 30-1010-01 Zero DAS data detect See Long Description for Details. Error: 186518 code 30-1010-01 Zero DAS data detect error detected by das das error: 1 code: 30-1010-01 name: zero detect. class: error description: zero das data was detected. potential cause / fru: 1. cam 2. cif 3. ddp Thu Mar 8 13:41:34 2007 186518 1 ct01 tgp_out tgp_errors.c 312 DAS Error. Code : 30-1010-01 Error detected by : CIF Zero DAS data was detected. EN 184 SR 184
Applies to
- Models
- HiSpeed
- Product
- HiSpeed YMS Series HiSpeed Series DXi, FXi, LXi, ZXI, and NXI Including Single / Twin detector and Glass / Metal Tubes. See Long Description for Details Subsystem/Components DXI (NP) FXI,LXI (NP+) ZXI (NP++ NXI (NP+ Twin NXI (NP++ Twin OC O2 running Unix or upgraded to Linux PC Gantry Alighnment Lights Halogen Laser Laser Laser Laser Mechanics Common Electronics firmware is different twin Det/Das Twin Det/DAS Table IMS no Yes Yes Yes Yes Generator Jedi Jedi NP++ Jedi Jedi NP++ Jedi Tube D3142T D3142T D3152T D3142T D3152T PDU Common Options List for V/R 5.5, 6, 7, 9, and 10 System 3D Power 440ma Max Image Disk 20000 Smart Prep Fast Scam 0.7 sec Performix 530 cooling Smart Recon Fast Scan 0.8 sec Asymmetric Scan SmartView Fast Scan 1.0 sec Cardiac gating DemtaScan Helical 30 Max Smart Addition Navigator Helical 60 Max Respiratory Gating Helical Plus Helical 120 Max Reformat Cpmmect {rp Raw Data 1000 max VolumeRendering2 ImageDosk Remote Tilt CT Colono Pioneer Mod Enhanced Cooling The following Options name changes Option (V5.5x,V6.0x,V7.x,V9.x,V10x) Option (V8.1x and V11.x) 3D >> Volume Analysis2 Navigator >> Navigator2,VolumeRendering2,VolumeAnalysis2 CtPerfusion >> CTPersusion2 Reformat >> Functool Options V8.1x and V11.x VolumeAnalysis2 Power 440ma ImageDisk 2000 Helical Plus Smart Prep Fast Scan 0.7 sec Performix 530 cooling Connect Pro Smart Recon Fast Scan 0.8 sec Asymmetric Scan ImageDisk SmartView Fast Scan 1.0 sec Cardiac Gating Pioneer MOD* DentaScan Helical 30 Smart Addition Power 200 ma max Navigator2/VolumeRendering2/ VolumeAnalysis Helical 60 Respiratory Gating Power 250 ma Max Power 300 ma Max Helical 120 Thickness 0.5mm CTColon Power 350 ma Max Raw data 1000 CtPerfusion2 AdvVesselAnalysis Remote Tilt Extended FOV Functool Enhance Cooling Performix Tube VolumeRendering2
- Component
- 2201229 - 12V DAS Power Supply used on HiSpeed YMS See LD for FEMC info 2201229 - 12V DAS Power Supply used on HiSpeed YMS Part #: 2201229 Part Type: GP CONSUMABLES Part Weight: 0.44 LBR Repairable: No Returnable: No Hazardous Material: N Description: POWER SUPPLY FRU: Yes Status: Orderable Comments: FLS - POWER SUPPLY 12V U/O CT HISPEED-YMS GANTRY M3 FW20 98
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required